| Catalog Number |
ES2816 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed and replaced as the implant was undersized and there was bad pump placement.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed and replaced as the implant was undersized and there was bad pump placement.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump, two cylinders and reservoir were received for evaluation.As examination of the components may not conclusively confirm or disprove the report of sizing issue and malposition, quality accepts the physician¿s observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|