• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE CD 12 MM THICKNESS; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE CD 12 MM THICKNESS; INSTRUMENT, KNEE Back to Search Results
Catalog Number 42527900302
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in progress.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a tibial articular surface provisional shim instrument was found to be disassembled and missing ball bearings during the sterilization process on an unknown date.There was no patient involvement and no adverse events have been reported as a result of the malfunction.All available information has been reported.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h11.Visual evaluation of the returned products identified that the products exhibit signs of use (nicked or gouged) and one or both of the ball bearings are missing (missing ball bearings are not returned).Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Complaint sample was evaluated and the reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE CD 12 MM THICKNESS
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21091797
Report Number0001822565-2025-00062
Device Sequence Number18428223
Product Code MBH
UDI-Device Identifier00889024245761
UDI-Public(01)00889024245761(10)63785355
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42527900302
Device Lot Number63785355
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2024
Supplement Date Received by Manufacturer05/01/2025
Initial Report FDA Received Date01/07/2025
Supplement Report FDA Received Date05/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-