| Catalog Number |
366408 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during use of bd vacutainer® no additive (z) plus tubes, an unspecified number of tubes had loose caps.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd had not received any samples for investigation.A total of 29 retained samples were sent for functional tests, and all tests showed no stopper creepout or stopper pop off.Based on the investigation results, no root cause from the manufacturing process has been identified as a contributor to the reported defect of stopper creepout.Based on a review of the device history record for the incident lots, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: stopper creepout.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported during use of bd vacutainer® no additive (z) plus tubes, an unspecified number of tubes had loose caps.There was no health impact or consequences reported.
|
| |
|
Search Alerts/Recalls
|