• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES; CONTAINER, SPECIMEN, STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES; CONTAINER, SPECIMEN, STERILE Back to Search Results
Catalog Number 366408
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during use of bd vacutainer® no additive (z) plus tubes, an unspecified number of tubes had loose caps.There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
E1.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary bd had not received any samples for investigation.A total of 29 retained samples were sent for functional tests, and all tests showed no stopper creepout or stopper pop off.Based on the investigation results, no root cause from the manufacturing process has been identified as a contributor to the reported defect of stopper creepout.Based on a review of the device history record for the incident lots, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: stopper creepout.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Event or Problem Description
It was reported during use of bd vacutainer® no additive (z) plus tubes, an unspecified number of tubes had loose caps.There was no health impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES
Common Device Name
CONTAINER, SPECIMEN, STERILE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key21092750
Report Number1024879-2024-01349
Device Sequence Number21066504
Product Code FMH
UDI-Device Identifier30382903664086
UDI-Public(01)30382903664086(11)240522(17)250930(10)4143690
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2025
Device Catalogue Number366408
Device Lot Number4143690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2024
Supplement Date Received by Manufacturer02/27/2025
Initial Report FDA Received Date01/07/2025
Supplement Report FDA Received Date03/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-