Model Number 1553201080
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient implanted with rods having l3, s1 free hands technique for open surgery with solera screw fenestrated for stenosis lumbar with osteoporosis.It was reported that rods were broken.Revision surgery was performed to place new rods.There was no patient symptoms reported.Patient had recovered from the ailments.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Radiographic image review results: 2 panel x-ray ap/ lateral l3-s1 fusion with cement augmentation, the rod on one side appears fractured between l4-5 just seen on ap image.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis # (b)(4): part # 1553201080, lot # 0561122w visual optical inspection revealed some wear and indentions on it from the seating and tightening of the set screw.Macroscopic inspection of the fracture surface revealed a flat fracture surface but no signs of cyclic fatigue.The rods appear to be broken from overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDTRONIC MEDICAL COSTA RICA S.R.L.
coyol free zone building b7
multi-tenant 7-8
coyol 20113
CS  20113
Manufacturer (Section G)
MEDTRONIC MEDICAL COSTA RICA S.R.L.
coyol free zone building b7
multi-tenant 7-8
coyol 20113
CS   20113
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21093209
Report Number3014725904-2025-00006
Device Sequence Number21026608
Product Code KWP
UDI-Device Identifier00643169015883
UDI-Public00643169015883
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1553201080
Device Catalogue Number1553201080
Device Lot Number0561122W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2024
Supplement Date Received by Manufacturer01/21/2025
04/14/2025
Initial Report FDA Received Date01/07/2025
Supplement Report FDA Received Date01/28/2025
04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-