| Model Number |
1845010 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis found that not able to confirm reported fault as difficult to insert tool into the attachment.Bearings were broken and corroded.Not able to perform pre repair temperature test as bearing found corroded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to procedure indigo attachment had heating issue.Handpiece was heating within one minute.Device was running on the speed of 60000rpm in forward mode and manual irrigation was used.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|