| Model Number |
LX3 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that before cataract surgery, the lower part of the optical head of an ophthalmic microscope was detached.The details of the procedure was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections d.9, h.3, h.6 and h.11.The company representative was able to confirm the reported event.The optical head was replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance based review of the serial number was performed and a potential contributing factor to the reported complaint was identified.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is attributed to the nonconforming optical head.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident/event data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|