| Model Number |
376297 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Inflation Problem (1310); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The passeo-18 peripheral balloon catheter was selected for treatment of a lesion with 75 percent stenosis degree in a mildly tortuous part of the distal infrapopliteal artery.The balloon was used for pre-dilatation, but it could not inflate.The device was withdrawn, and a balloon leakage was found.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation confirmed that the balloon has been inflated.Upon inflation during functional testing, a fine jet of water was seen to emerge from the proximal balloon portion.Microscopic inspection revealed a small pinhole in the balloon about 2 mm distal to the proximal radiopaque marker.Deep scratches were observed in close vicinity to the pinhole which have likely been caused by a hard, sharp-edged object such as e.G., anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause was determined.The most probable root cause for the reported event is related to the patient's anatomy.
|
| |
|
Search Alerts/Recalls
|