• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE Back to Search Results
Model Number 2700TFX
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: additional manufacturer narrative: due diligence attempts have been made to obtain additional information / the status of the explanted device for return.At this time, additional information / the device has not been provided.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through the implant patient registry that a patient with a 25mm 2700tfx aortic valve, was explanted after an implant duration of eight (8) years, two (2) months due to unknown reason.The explanted valve was replaced with a 25mm 11500a valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: due diligence attempts were made to gather additional information regarding a device failure mode, however additional information is not available at this time.Patient medical records were not provided by the hospital for review, or the medical records provided do not clarify the device failure mode.In this case, there is no evidence of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Per the instructions for use (ifu), reoperation/explant are known potential adverse events associated with bioprosthetic cardiac valves and annuloplasty rings.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, a definitive root cause cannot be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key21094035
Report Number2015691-2025-00114
Device Sequence Number18428240
Product Code DYE
UDI-Device Identifier00690103194043
UDI-Public(01)00690103194043(17)200105
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/05/2020
Device Model Number2700TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2024
Supplement Date Received by Manufacturer01/29/2025
Initial Report FDA Received Date01/07/2025
Supplement Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age56 YR
Patient SexMale
-
-