MAUDE Adverse Event Report: EXACTECH, INC. CR TIBIAL INSERT SZ3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CR TIBIAL INSERT SZ3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-23-11
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Synovitis (2094)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report number: 1038671-2023-00782.D10 concomitants: (b)(6) 200-02-38 - three peg patella 38mm.(b)(6) 204-04-33 - trapezoid tibial tray sz 3f/3t.(b)(6) 204-32-01 - fluted stem extension 11l x 12 mm.(b)(6) 204-70-00 - tibial stem ext.Sc.(b)(6) 200-01-03 cemented cr femoral sz 3.The product associated with the reported event is within the scope of recall z-0019-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00782.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported by legal notification, the patient had an initial left tka on (b)(6) 2014.The patient was revised approximately 8 years and 2 months later on (b)(6) 2022.Joint fluid was obtained and sent for culture and sensitivity x3.Extensive synovitis from the polyethylene was noted.The femoral component was grossly loose and was removed with osteotomes.The tibial component and patella were also grossly loose and were removed.The patient tolerated the procedure well.No other patient information/medical history reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CR TIBIAL INSERT SZ3, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
MDR Report Key21095806
Report Number1038671-2025-00131
Device Sequence Number15732871
Product Code JWH
UDI-Device Identifier10885862040251
UDI-Public10885862040251
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2015
Device Catalogue Number200-23-11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Supplement Date Received by Manufacturer01/08/2025
Initial Report FDA Received Date01/07/2025
Supplement Report FDA Received Date01/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexUnknown
-
-