| Catalog Number |
286210110 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the following issues were observed following the use of the ancillary set: difficulty using the reducers because the screwdrivers did not fit in the reduction screwdrivers.The screwdrivers did not penetrate fully into the reducer screwdrivers, so it was not possible to insert the plug onto the implant.In addition, the rod stabilizer did not allow it to perform its function.The end was covered with a rubber cap with a hole for the screwdriver; this cap could not be removed and did not allow the stabilizer to be positioned on the rod.
|
| |
|
Event or Problem Description
|
|
Additional information received states that the patient outcome was reported to be fine and the procedure was completed successfully.There was no additional medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: b5.Corrected: h6 (health effect, impact code & health effect, clinical code).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.¿.Device history review (dhr): a review of the receiving inspection of single innie inserter was conducted identifying that lot number nw130810 released in four batches.Supplier: (b)(4).Batch 1: lot qty of (b)(4) units were released on 02 feb 2013 with no discrepancies.Batch 2: lot qty (b)(4) units were released on 11 jan 2013 with no discrepancies.Batch 3: lot qty of (b)(4) units were released on 11 jan 2013 with no discrepancies.Batch 4: lot qty of (b)(4) units were released on 13 sep 2016 with no discrepancies.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|