ETHICON INC. PDSII CLR 18IN 4-0 S/A P-3 PRM MP; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number Z494G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: when did the suture break (in the package, during removal from the package, during handling prior to using on the patient, or during use on the patient).Please provide details for each of the affected devices.During use on the patient.What is the procedure name? unk.
 
Event or Problem Description
It was reported that a patient underwent an unknown surgery on (b)(6) 2024 and suture was used.The suture broke during the operation.There were no adverse consequences to the patient.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 2/10/2025 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d4, d9, h3, h4, h6 a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One representative unopened sample was received for analysis.The complaint sample was not received for evaluation.Upon initial inspection of the sample, no external damages were observed on the external packet.In order to evaluate the condition of the returned sample, the packet was opened.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII CLR 18IN 4-0 S/A P-3 PRM MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21101083
Report Number2210968-2025-00280
Device Sequence Number21122367
Product Code NEW
UDI-Device Identifier10705031061170
UDI-Public(01)10705031061170(17)280731(10)TJMHZC
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberZ494G
Device Lot NumberTJMHZC
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2024
Supplement Date Received by Manufacturer02/10/2025
Initial Report FDA Received Date01/08/2025
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-