| Catalog Number |
CAR-02400 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found the lure hub/fitting has crack during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The customer returned one connector assembly for analysis.Signs of use were observed.Visual analysis revealed the red arterial luer hub contained a crack.Stress marks/scratches were also noted on both hubs.Microscopic examination confirmed the damage.The appearance of this damage is consistent with overtightening or repeated tightening of the hubs.No defects or anomalies were observed on any other portion of the returned device.Luer connections were functionally tested per the product instructions for use (ifu).The ifu provided with this kit instructs the user, "establish and maintain catheter patency.Solution and frequency of flushing a venous access catheter should be established in hospital/institutional policy".A lab inventory syringe filled with water was used to flush each extension line.When flushing the red arterial extension line, water was observed leaking from the crack.No leaking was observed from the blue venous luer hub.A manual tug test confirmed both luer hub were securely attached to their respective lumens.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure".The report of a cracked luer hub was confirmed through complaint investigation of the returned sample.Visual analysis revealed that the red arterial luer hub contained a crack and damage consistent with overtightening or repeated tightening of the hubs.A device history record review was performed, and no relevant findings were identified.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found the lure hub/fitting has crack during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.The associated emdr numbers are 9680794-2025-00030, 9680794-2025-00031, 9680794-2025-00032, 9680794-2025-00029.
|
| |
|
Search Alerts/Recalls
|