Model Number EVOLUTFX-23
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that during the attempted implant of this transcatheter bioprosthetic valve, the valve would not fully expand.The valve was withdrawn from the patient and deployed in warm saline, however, the valve continued to not fully expand.A different valve was used to complete the procedure.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.Product analysis: upon receipt of the suspect complaint device at medtronic¿s quality laboratory, visual analysis showed the device was received via fedex inside its original packaging and jar filled with clear unknown solution.The valve skirt appeared to be in partially crimped state.All leaflets were flexible and appeared intact.Coaptation: all leaflets appeared partially twisted in the closed position.All commissures were intact.The valve was submerged in warm saline to reset valve frame.The valve was then placed in a cold saline bath to perform loading simulation per instructions for use.Upon loading, both frame paddles appeared seated within the paddle attachment pockets.The capsule was advanced covering the frame paddles and outflow crown struts.The capsule was advanced until the capsule guide tube covered the commissure pad of the valve.The inflow cone was advanced to crimp the inflow portion of the valve.The capsule was fully advanced over the valve; no visual or tactile anomalies were noted.To simulate the reported event, the valve was deployed in a warm saline bath.The deployment knob was rotated to expose the valve.The valve continued to be deployed up to the point of no recapture; no anomalies were noted.The valve was fully deployed and appeared to exhibit a complete dilation; no anomalies were observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key21102141
Report Number2025587-2025-00170
Device Sequence Number18430471
Product Code NPT
UDI-Device Identifier00763000370688
UDI-Public00763000370688
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberEVOLUTFX-23
Device Catalogue NumberEVOLUTFX-23
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2022
Initial Report FDA Received Date01/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-