Material#: 515003 batch number#: unknown verbatim#: hope you had a wonderful christmas, over last 2 weeks we have 3 reported cases of leak at the connection point between icu medical tubing and phaseal injector luer lock.Hence wanted to check with you if this is something you may have heard from any other site as well.Has happened over the past few weeks--as i was made aware today.Apparently the third time was today and prompted the email to me.Based on customer response: ¿ are the pumps the icu cadd, and icu cadd sets? yes ¿ also, are the sets, fixed male luer, or a spinning collar/male swivel luer? fixed male luer ¿ the date this happened?: 12/26/24 ¿ while being used on a patient? yes ¿ what happened¿ the connection got unscrewed and extension set was disconnected ¿ what type of procedure? chemo infusion ¿ product: do you have the affected two items that leaked? were they discarded? items were discarded ¿ lot # can you please provide? and would you happen to have the packaging as well? or is the attached picture the representative lot # from the pars? attached pictures are only thing we have ¿ can you please provide the expiration date-please? not available ¿ was there any patient harm?.No immediate harm noticed, patient exposed to the 5fu he was receiving.After he manually connected the points again he went to ed to get it checked, where port was found to be functional and patient was discharged ¿ any pictures? attached earlier ¿ what is the patient outcome? patient experienced wetness, upon checking he found the dislodged connection, he thinks this lasted for about 30 minutes.He connected the points himself, went to ed to make sure and infusion was continued.The pump was unhooked after completion of infusion at infusion center later in the day.¿ are there any adverse events/serious injuries? not reported ¿ was there a delay of, or change in, the course of treatment due to the event? no ¿ what type of procedure is being performed? 5fu-pump for at home infusion.
|
|
(b)(4).Follow up mdr for device evaluation: no photos or physical samples that display the reported condition were available for investigation.Product undergoes inspections throughout the manufacturing process, including testing to verify all critical dimensions are within specification.As the lot involved in this incident is unknown, a device history review could not be performed.Based on the available information we are not able to identify a root cause related to our device or manufacturing process at this time.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
|