|
One pe074f5 swan ganz bipolar pacing was returned for evaluation.Customer report of pin connectors should have been labeled as distal and proximal respectively, but both were found incorrectly labeled as proximal was confirmed.As received, the labels on both of the pin connectors were proximal.No visible damage was observed from the catheter body, balloon, windings and returned syringe.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 minutes without leakage.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The device history record review could not be completed since the lot number is unknown.As part of the manufacturing process, 100% of the units go through a visual inspection as per procedures ensamblajes de bipolar pacing catheter assembly, and final inspection for electrode and tip.This inspection includes to verify the identification in each pin plug.Japanese ifu indicates instructions to handle the catheter with caution for proper functioning of the pacing catheter.However, there is no indications to verify for pins identification.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint and a capa has been generated to perform an investigation and corrective preventive actions for this issue.
|