The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device in an off-label location, not performing an x-ray immediately after the procedure to confirm placement, inadequate fixation, the patient undergoing an mri, implanting the neurostimulator after the expiration date, not prepping the skin with an antiseptic solution, not irrigating the incision site with an antibiotic solution before closure, improperly closing the surgical site, multiple tunneling attempts, using inappropriate tools, and not prescribing antibiotics pre-operatively have been ruled out as potential causes for the reported issues.Additionally, the patient having contraindicating conditions has been ruled out.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of impaired wound healing and erosion is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
|