MAUDE Adverse Event Report: CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number STQ4-RCV-A0, STQ4-SPR-B0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erosion (1750); Impaired Healing (2378)
Date of Event 12/17/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported impaired wound healing and erosion.Antibiotics were prescribed and an explant procedure was performed in (b)(6) 2024.However, an exact date is unknown.The patient was implanted with two new neurostimulators on (b)(6) 2024 and the patient is doing well.
 
Additional Manufacturer Narrative
The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device in an off-label location, not performing an x-ray immediately after the procedure to confirm placement, inadequate fixation, the patient undergoing an mri, implanting the neurostimulator after the expiration date, not prepping the skin with an antiseptic solution, not irrigating the incision site with an antibiotic solution before closure, improperly closing the surgical site, multiple tunneling attempts, using inappropriate tools, and not prescribing antibiotics pre-operatively have been ruled out as potential causes for the reported issues.Additionally, the patient having contraindicating conditions has been ruled out.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of impaired wound healing and erosion is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM PERIPHERAL NERVE STIMULATOR
Common Device Name
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central blvd s
pompano beach FL 33064
Manufacturer Contact
tena jimmerson
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key21106149
Report Number3010676138-2025-00005
Device Sequence Number18332388
Product Code GZF
UDI-Device Identifier00818225020464
UDI-Public(01)00818225020464(17)221001(01)00818225020495(17)221201(21)2B12590-108
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K171366
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/01/2022
Device Model NumberSTQ4-RCV-A0, STQ4-SPR-B0
Device Lot Number2B12369-53, 2B12590-108
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2024
Initial Report FDA Received Date01/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-