SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 10MM X 30MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72204409 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
12/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy acl reconstruction surgery, when the screwdriver was removed, the proximal half of the biosure screw came out with the driver.The biosure screw was broken in half.The distal part of the biosure regenesorb was left in the patient¿s tibia, the proximal part of the screw was removed with tweezer from patient.The procedure was completed with another surgical technique; the distal broken part of the screw held the graft in tibia tunnel with fair biting force and the surgeon reinforced the holding strength by securing the remaining graft and sutured with a cortical screw and washer on a separate site more distal to the tibia tunnel end.There was a surgical delay of less than 30 minutes, and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the reported device was not returned to the designated complaint unit for independent evaluation.However an image evaluation was performed and found two pictures of the label of the device.Another picture showed the device being held in a hand with the tip broken off.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification, found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
|
| |
|