• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number L-1742K
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during an open ophthalmic procedure, in ocular oblique stitching while employing a continuous suturing te chnique, the thread broke for the two sutures.To resolve the issue, a new suture was used.There was no patient injury.
 
Additional Manufacturer Narrative
D10.Concomitant medical product: l-1742k polysorb* 6-0 vio 30cm ss24da (lot d4a2681y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted two sutures with needles attached and observed in an open foil pouch.The sutures appeared to be intact.The length of the sutures present in the image could not be evaluated.It was reported that the thread broke.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21106838
Report Number9612501-2025-00102
Device Sequence Number8023183
Product Code GAM
UDI-Device Identifier10884521046245
UDI-Public10884521046245
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL-1742K
Device Catalogue NumberL-1742K
Device Lot NumberD4A2681Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer03/24/2025
Initial Report FDA Received Date01/08/2025
Supplement Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-