| Catalog Number |
89-8522-450-60 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Overheating of Device (1437); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: spain.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the adapter loosened from the handpiece during operation, then overheated and stopped working.No consequences or impact to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following section was corrected: d4 primary unique device identification (udi) number.No product was returned.Review of the provided pictures found the attachment and signs of wear; however, product evaluation is needed to determine if the device is overheating and functioning properly.Review of the device history record identified no deviations or anomalies during manufacturing.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|