| Catalog Number |
TSVB10 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided.G4: premarket identification: k011245|k002188.
|
| |
|
Event or Problem Description
|
|
It was reported a fracture of the upper part of the implant at tooth site 15 observed when the clinical procedure was finished and patient came to the clinic with the implant fractured.The process was completed with another implant.
|
| |
|
Event or Problem Description
|
|
It was reported a fracture of the upper part of the implant at tooth site 15 observed when the clinical procedure was finished and patient came to the clinic with the implant fractured.The process was completed with another implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D1: brand name unknown / not provided.D4: additional device information updated to tsvb10 unknown lot number.D4: unique identifier (udi) number not available.G3: date received by manufacturer.G4: premarket identification k011028/k013227.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).Zimvie received one (1) tsvb10, (impl tapered scr-v sbm 3.7mm 3.5mm 10mm), for evaluation.Visual evaluation was performed; a fracture has been identified on the implants collar.A third-party removal tool attached, and was not identified as reported did not contribute to the event.Dhr review and complaint history review by lot number could not be performed, as the lot number associated with the reported product is not available.However, zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.A complaint history review by item number was conducted for the [tsvb10] dating back to 12 months from now.The complaint history review revealed that there are no existing non-conformances/capa/hhe/d/product holds for the reported product for similar event.Review completed utilizing keywords: dental: functional: fracture: implant.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was clinician selects implant that is unable to resist long-term occlusal loading.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the collar.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Search Alerts/Recalls
|