• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 7 10MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 7 10MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151650710
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 12/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient was being revised due to complaint of pain.The primary components were an attune size 7 ps femur, rotating platform tibial base plate, a size 7 x 10 mm insert and 32mm anatomic patella.During the case, it was discovered that the patella component was loose.The surgeon removed the patella component, cleaned off the bone re-drilled the peg holes and implanted a 32 mm anatomic patella.He also trialed a thicker tibial insert to address instability and ultimately implanted a 14 mm size 7 insert.Unknown surgical delay.Doi: unknown.Dor: (b)(6) 2024.Affected side: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS RP INSRT SZ 7 10MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21110718
Report Number1818910-2025-00484
Device Sequence Number7463501
Product Code NJL
UDI-Device Identifier10603295052746
UDI-Public(01)10603295052746
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151650710
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/16/2024
Initial Report FDA Received Date01/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient Weight118 KG
-
-