| Model Number |
DISSECTOR, CURVED 4.0MM X 13CM |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the shaver left residues during activation.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a different device.It was not necessary to switch the surgical technique or do a second surgery.On 09-oct-2023, update dw further information were provided that the customer tried to remove the debris but was not 100% sure if everything could been retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to prying/leveraging the attachment during use.
|
| |
|