Model Number T10 HEXALOBE, CMP FT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 12/13/2024, it was reported by a sales representative via sems-06734836 that (qty.3) of an ar-8737-38 driver shaft, t10 hex, cannulated broke when attempted to remove three 3.5 fully threaded headless compression screws in the midfoot region while using proper technique.They tried to use the arthrex mini trephine, and after many attempts, they used a non-arthrex product to core around the headless compression screws to remove them.No pieces broke off inside the patient reported.The case was completed successfully.This was discovered during a midfoot hardware removal procedure on (b)(6) 2024, with no reported patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during removal of the screws.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T10 HEXALOBE, CMP FT
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21113491
Report Number1220246-2025-09312
Device Sequence Number18044961
Product Code LXH
UDI-Device Identifier00888867127166
UDI-Public00888867127166
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberT10 HEXALOBE, CMP FT
Device Catalogue NumberAR-8737-38
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2024
Supplement Date Received by Manufacturer06/26/2025
Initial Report FDA Received Date01/09/2025
Supplement Report FDA Received Date07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-