| Model Number |
413997 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
This lead was explanted due to dislodgement.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual, mechanical and electrical inspection.The visual analysis revealed a cut into the insulation 11,5 cm distal to the is-1 connector pin, which most likely resulted from the explantation procedure.However, a thorough analysis of the lead did not show any deviations which might have led to the reported dislodgement.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|
| |
|
Event or Problem Description
|
|
This lead was explanted due to dislodgement.Should additional information be received, this file will be updated.
|
| |
|
Search Alerts/Recalls
|