| Model Number |
GALILEO® CAPTURING ROD NUT |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 12/12/2024, it was reported by a sales representative via (b)(4) that a 5033-100 galileo® capturing rod and an 5046-100 galileo® capturing rod nut will not unscrew and are stuck in the 5030-000 galileo® lag screw inserter.A backup set was available, and the case was completed without adverse effects on the patient.This was discovered during a procedure, with no reported patient harm.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged 5046-100 serial/batch number (b)(6) was received for evaluation.The device arrived stick to the 5033-100 and 5030-000.Functional testing was not performed because the devices arrived stuck together for investigation.However, attempts to unscrew the knob were unsuccessful.A visual inspection revealed damaged on the threads of the knob.The most likely cause of the reported failure is misuse by the end user, who most likely cross-threaded the devices.
|
| |
|