• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE Back to Search Results
Model Number M00553680
Medical Device Problem Codes Positioning Failure (1158); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a15 captures the reportable investigation finding of stent partially deployed.Block h11: an axios stent and electrocautery enhanced delivery system was received for analysis.Visual examination of the returned device found the stent partially deployed.Functional inspection was performed by sliding the catheter lock to the right to unlock the catheter, then the catheter control hub was moved up and down.The catheter passed through the luer without resistance.Also, the stent hub was moved down to the second and fourth position, and the stent was deployed and presented slow expansion.No other problems were noted with the stent and delivery system.Product analysis confirmed the reported event of stent failure to deploy as the device was received not fully deployed.The observed problem of stent partially deployed is known and documented in the labeling; therefore, a review and analysis of all available information indicated the most probable cause is known inherent risk of device.A labeling review was performed and, from the information available, there is no information that this device was used in a manner inconsistent with the instructions for use (ifu)/ product label.Additionally, stent partially deployed is noted within the ifu as possible adverse event associated with the use of the device.
 
Event or Problem Description
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was to be implanted to treat a cyst during a cystogastronomy procedure performed on (b)(6) 2024.During the procedure, the axios stent would not deploy.There were no patient complications reported as a result of this event.Note: this complaint has been deemed mdr reportable event based on the investigation result which revealed that stent was partially deployed.Please see block h11 for the full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Common Device Name
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key21114448
Report Number3005099803-2024-06811
Device Sequence Number15338197
Product Code PCU
UDI-Device Identifier00191506008086
UDI-Public00191506008086
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K233318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/24/2024
Device Model NumberM00553680
Device Catalogue NumberM00553680
Device Lot Number0031891875
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2024
Initial Report FDA Received Date01/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-