| Model Number |
3300 |
| Medical Device Problem Codes |
Failure to Charge (1085); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Visual inspection of the device identified no abnormalities or visual inspection of the device revealed external damage consistent with the programmer being dropped.The programmer was powered on and it was confirmed that the touchscreen was intermittently unresponsive due to not detecting touch inputs.No error or fault codes were recorded in the programmer's logfile.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
|
| |
|
Event or Problem Description
|
|
It was reported that this integrated programmer no longer charging.The field representative tested different power cords and, lacked an alternative battery for testing, attempted removing and reinserting the existing battery.The field representative completed a device check but remained uncertain about the device's reliability.The representative's manager had already requested a replacement.This programmer was returned.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|