• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND GLENOSPHERE 40+8; SHOULDER IMPLANT - GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND GLENOSPHERE 40+8; SHOULDER IMPLANT - GLENOSPHERE Back to Search Results
Catalog Number 550540008
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had their original surgery (b)(6) 2024.While at the gym throwing around a 15lb weight, patient felt a pop in his shoulder.The surgeon brought patient in on (b)(6) 2024 to revise his left shoulder because the glenosphere had disassociated from the baseplate.After removing this implant, surgeon noticed that the safety screw housed within the glenosphere still seemed to be set into the glenosphere and the threads never engaged the baseplate.Surgeon removed this implant, and placed a new one with no issue.The taper engaged, and the safety screw was screwed into the baseplate with no incident.There was no surgical delay.Doi: unknown dor: (b)(6) 2024 affected side: unknown shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Additional information received states that there was no surgical delay.The affected side was the left shoulder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOSPHERE 40+8
Common Device Name
SHOULDER IMPLANT - GLENOSPHERE
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21115294
Report Number1818910-2025-00565
Device Sequence Number8732808
Product Code PHX
UDI-Device Identifier10603295547174
UDI-Public(01)10603295547174(17)270430(10)332931
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K212737
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number550540008
Device Lot Number332931
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/23/2024
Supplement Date Received by Manufacturer01/13/2025
Initial Report FDA Received Date01/09/2025
Supplement Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-