Catalog Number 231818
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a ligament reconstruction surgical procedure the milagro advance screw device anchor was broken off.All broken parts were removed.Another like device was used to complete the procedure.It was unknown if there was delay to the procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d9, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The product was returned to j&j medtech orthopaedics for evaluation.J&j medtech orthopaedics then conducted visual inspection of the device received.Visual inspection revealed that the anchor was returned without structural anomalies.The overall complaint was unconfirmed as the observed condition of the milagro advance screw 9x23mm would not contribute to the complained device issue.Based on the investigation findings and since no broken areas were identified, it was conclude that there is not enough evidence to adjudicate a root cause for the issue experienced by the customer.Therefore, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MILAGRO ADVANCE SCREW 9X23MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key21117185
Report Number1221934-2025-00157
Device Sequence Number20698789
Product Code MAI
UDI-Device Identifier10886705022342
UDI-Public(01)10886705022342(10)275L341(17)270228
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K143660
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number231818
Device Lot Number275L341
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2025
Supplement Date Received by Manufacturer01/27/2025
01/31/2025
Initial Report FDA Received Date01/09/2025
Supplement Report FDA Received Date01/31/2025
02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-