• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; BREATHING CIRCUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; BREATHING CIRCUIT Back to Search Results
Model Number RT205
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Section g4: the mr850gju respiratory humidifier is not sold in the usa, but is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Fisher & paykel (f&p) healthcare is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A healthcare facility in china reported that an mr290v vented autofeed humidification chamber provided with an rt205 adult breathing circuit was found leaking during use.There was no patient harm reported.
 
Additional Manufacturer Narrative
(b)(4).Correction: section h11: amended comment for section g4.Section g4: the rt205 adult breathing circuit is not sold in the usa, but is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Section h11: method: the subject mr290v autofeed humidification chamber provided with the rt205 adult breathing circuit, additional information and photographs were requested to be provided to fisher & paykel (f&p) healthcare for analysis.Results: the subject device and photographs could not be returned to f&p healthcare as it had been discarded by the healthcare facility.Conclusion: based on the limited information provided by the healthcare facility and without the return of the subject device, we are unable to confirm the reported issue or determine the cause of the reported event.Every mr290v vented autofeed humidification chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v vented autofeed humidification chamber would have met the required specification at the time of production.The user instructions for the rt205 adult breathing circuit state the following: - do not use the chamber if the seals are not intact when received, or if it has been dropped.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- ensure there is a water supply connected to the chamber and that water is present within the chamber.- appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.- ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.- visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.
 
Event or Problem Description
A healthcare facility in china reported that an mr290v vented autofeed humidification chamber provided with an rt205 adult breathing circuit was found leaking during use.There was no patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
BREATHING CIRCUIT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key21121110
Report Number9611451-2025-00003
Device Sequence Number8683150
Product Code BTT
UDI-Device Identifier09420012430410
UDI-Public(01)09420012430410(10)2103163077(11)240530(17)290530
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
SEE H11
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT205
Device Catalogue NumberRT205
Device Lot Number2103163077
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2024
Supplement Date Received by Manufacturer01/14/2025
Initial Report FDA Received Date01/09/2025
Supplement Report FDA Received Date02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PHILIPS V60 VENTILATOR
Patient SexUnknown
-
-