|
This report has been identified as b.Braun internal report number (b)(4).Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.Underdose general information: complaint: (b)(4).Information to the sample: model: perfusor space article number: 8713030 serial number/batch: (b)(6) software version: 688n030005 hours of operation: 2579 further information: n/a investigation results: history inspection: the device history files were read out and analyzed.The customer specified 2024-12-10 as the date of the incident.The last therapy on this day was analyzed.2024-12-10 at 10:45 a bd plastipak 50ml was selected.The syringe was filled to approx.49ml.The drug short name morphmt was selected.The therapy was started with a flow rate of 0 ,19ml/h at 10:45.The therapy was terminated by the user after 3 hours and 37 minutes.0.73ml were infused in total.No abnormalities can be found in the device history files.Visual inspection: a visual inspection was performed.The cover caps on the screw pillars, and the production seal on the lower housing were intact and undamaged.No visible damaged parts are to locate.Functional inspection: a functional test was performed.The device passes the self-test.A bd 50 ml syringe was inserted, and the pump identified the syringe, and it could be selected from the menu.It was possible to put the pump in operation.Individual inspection: a delivery accuracy measurement according to iec 60601-2-24 was arranged.Here a nominal flow rate of 5 ml/h was chosen.The assessed mean deviation "a" of the second operating hour was measured and resulted in a value of + 0.29%.Accuracy of set delivery rate should be ± 2 % according to iec/en 60601-2-24.Disassembling: during the investigation no faults could be detected, to investigate the inside of the device, only the upper housing was removed.No damage or soiling could be found.Judgment: the complaint could not be confirmed.Summing up all tests, the perfusor space operated within our specification.No product deviation.
|