|
Additional information: d9, g3, h3, h6.The complaint allegation was not confirmed.One unpackaged ar-1923bc-1, serial/batch number (b)(6), was received for evaluation.The device was received without the bio/screw, and no evidence of breakage was received.Visual inspection revealed no damage to the shaft or the tip.However, the device is missing the bio and the eyelet.Functional testing was performed moving the inner shaft inside the outer looking for resistance.It was noted no issues.The most likely cause is attributed to misuse due to user error improver bone preparation, using excessive force to pry and/or leverage the device.Directions for use (dfu) section g.Precautions.4.Pushlock and swivelock suture anchors only: during anchor insertion, ensure that the angle of anchor insertion is coaxial to that of the previously prepared bone socket.
|