| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: work order search: no similar complaints within associated lots were found.Manufacturer narrative: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was later removed and replaced for a same size, different power lens and the problem was resolved.Cause of the event is unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product evaluation: lens was returned in liquid within a microcentrifuge vial.Visual inspection found no visable damage to the lens.Dimensional and functional inspection found lens to be manufactured within specifications.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h3: device evaluation: lens was returned in liquid within a microcentrifuge vial.Visual inspection found no visable damage to the lens with fiber on the lens.Dimensional and functional inspection found lens to be manufactured within specifications.Claim#: (b)(4).
|
| |
|
Search Alerts/Recalls
|