• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. STENT; LVIS EVO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. STENT; LVIS EVO Back to Search Results
Model Number LEV2517-PMA
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation findings: items returned: - n/a; the device remains implanted in the patient.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging were not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot further examine if a condition existed that would have contributed to the reported event.Without imaging, the investigation cannot further verify if the event occurred as described, nor could the investigation further evaluate the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part / lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential adverse events the following potential risks and complications associated with general anesthesia, cerebral angiography, intracranial catheterization, intracranial stent placement or intra-saccular coil deployment have been identified below: ¿ allergic reaction, including but not limited to: contrast dye, nitinol metal, and any other medications used during the procedure; ¿ aphasia ¿ blindness; ¿ cardiac arrhythmia; ¿ coil prolapsed or migration into normal vessel adjacent to aneurysm ¿ complications of arterial puncture including pain, local bleeding, local infection and injury to the artery, vein or adjacent nerves; ¿ cranial neuropathy; ¿ death; ¿ device fracture, migration or misplacement; ¿ dissection or perforation of the parent artery; ¿ headache; ¿ hemorrhage (i.E., intracerebral hemorrhage (ich), subarachnoid hemorrhage (sah), or retroperitoneal (or in other locations)); ¿ hemiplegia; ¿ hydrocephalus; ¿ infection; ¿ injury to normal vessel or tissue; ¿ ischemia; ¿ mass effect; ¿ myocardial infarction; ¿ neurological deficits; ¿ occlusion of non-target side branches; ¿ pseudo aneurysm formation; ¿ reactions to anti-platelet/anti-coagulant agents; ¿ reactions due to radiation exposure; ¿ reactions to anesthesia and related procedures; ¿ reactions to contrast agents; ¿ renal failure; ¿ aneurysm rupture; ¿ stenosis of stented segment; ¿ seizure; ¿ stent thrombosis; ¿ stroke or tia (transient ischemic attack); ¿ thromboembolic event (t/e); ¿ vasospasm; ¿ visual impairment.Warnings: should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis device with compatible microcatheters or occlusion balloons.If repeated friction is encountered during lvis device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis device in the parent vessel without fully retrieving the device.The lvis device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions exercise caution when crossing the deployed / detached lvis device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis device delivery or manipulation, withdraw the unit and select a new lvis device.18.Position the lvis device for deployment by aligning the lvis implant distal radiopaque end markers approximately 7 mm past the aneurysm neck.Note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis device is not recommended and may result in device elongation.19.If lvis device positioning is not satisfactory, the lvis device may be recaptured and repositioned if it is not fully deployed.The lvis device may be recaptured until the point where the proximal end of the lvis device markers is aligned 3 mm proximally with the microcatheter distal marker band (approximately 75% deployed).Caution: if resistance is felt while recapturing the lvis device, do not continue to recapture the device.Withdraw the microcatheter slightly to un-sheath the lvis device (without exceeding the recapture limit), and then attempt to recapture the lvis device.Caution: the lvis device must not be re-deployed more than three times.Note: the lvis device delivery wire should not be utilized as a guidewire.Do not torque the lvis device.A torque device should not be used.20.If lvis device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are sufficiently proximal to the aneurysm neck to ensure an adequate landing zone.The lvis device will expand, and total length may foreshorten up to 60% from its undeployed length (refer to table 1a/b/c) as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length on each side of the aneurysm neck or target location to ensure appropriate neck coverage.Warning: do not detach the lvis device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis device to maintain access through the lvis device.Remove and discard the delivery wire.Warning: the lvis device delivery wire should not be utilized as a guidewire.Do not torque the lvis device.A torque device should not be used.23.Use the guidewire and microcatheter to access the aneurysm through the lvis device cells.Warning: observe lvis device marker position during placement of the microcatheter into the aneurysm to ensure that the lvis device does not migrate or dislodge from its deployed position.24.After the microcatheter is positioned within the aneurysm, detachable coils may be delivered into the aneurysm according to conventional methods.Warning: observe lvis device marker position during the coiling procedure to ensure that the device does not migrate from its deployed position.25.After placing the last coil, verify that the lvis device has remained patent and properly positioned.Advance a guidewire to the microcatheter tip and carefully remove the microcatheter through the lvis device cells.Note: a microcatheter may be positioned into the aneurysm sac prior to delivery of the lvis device.The microcatheter will be supported by the lvis device during delivery of embolic coiling.After completing the coiling, the microcatheter should be carefully removed to avoid dislodging the lvis device.27.Caution: carefully watch the lvis device distal and proximal markers when passing through the deployed lvis device with embolic coiling microcatheters to avoid displacing the lvis device.Investigation conclusion: the physical device was not available for evaluation to investigate if a condition existed that would have contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.This information may be updated if additional information is provided at a later date.
 
Event or Problem Description
It was reported that the lvis evo could not be recaptured in the microcatheter.Stent / coil procedure (acomm).A stent / coil procedure was performed to treat an acomm aneurysm.The microcatheter was used in a "jailed" technique to deliver lvis evo.A separate microcatheter was used to deliver one coil.When delivering lvis evo, physician attempted to recapture partially deployed (approximately 1/3 to 1/2) stent to reposition the evo.The stent would not recapture back into the microcatheter.Stent was delivered due to not being recapturable.The patient is doing well and discharged from the hospital.
 
Additional Manufacturer Narrative
Investigation conclusion: the stent was implanted and the pusher wire was discarded; however, only the microcatheter used in the procedure was returned for evaluation.Since the reported stent was not returned, this investigation could not test for or assess the alleged re-sheathing issue.The returned microcatheter was found to be in good condition; furthermore, an in-house stent was able to advance through the returned microcatheter and re-sheathe without resistance during the investigation.
 
Event or Problem Description
Additional information received reported that the final placement of the stent was not in the desired location.The physician started his initial delivery of the lvis evo (approximately 1/3 to ½ of the lvis evo was deployed).After the physician delivered the initial 1/3 to ½ of the stent, he paused to check the position of the stent and did a visual estimation of where the stent would be positioned if fully deployed.The physician decided to recapture the stent in an effort to reposition the microcatheter and redeploy the stent in an adjusted location.The stent could not be recaptured after multiple attempts.Since the stent could not be recaptured, the physician decided to move forward and deliver the stent in the initial location.The stent remains implanted in the patient and the pusher wire was discarded; however, the microcatheter used in the procedure was returned for evaluation.Please see section h11 for investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STENT
Common Device Name
LVIS EVO
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key21126084
Report Number2032493-2025-00051
Device Sequence Number18021876
Product Code QCA
UDI-Device Identifier00842429117750
UDI-Public(01)00842429117750(11)231113(17)261031(10)0000457217
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P170013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLEV2517-PMA
Device Lot Number0000457217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer01/13/2025
Initial Report FDA Received Date01/10/2025
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER: STRYKER; HEADWAY 17 MICROCATHETER: LOT# 0000672418; TARGET XL COIL: STRYKER NEUROVASCULAR
Patient Age66 YR
Patient SexFemale
Patient Weight76 KG
-
-