| Catalog Number |
FEM12040 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Perforation (2205); Misfire (2532); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a stent graft placement procedure, the stent allegedly would not deploy.It was further reported that the tip of the catheter was allegedly damaged.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was returned for evaluation and one of the stent struts was found perforating the sheath while the stent was partially deployed.It is considered the perforation of the sheath led to the impossibility to completely deploy the stent graft which leads to confirmed results.Based on the provided information and the evaluation of the returned sample, the investigation is closed with confirmed results for sheath perforation and misfire of the stent graft is considered a cascading event.A definite root cause for the reported event could not be determined.The intended placement of the device in treating aneurysm is an off-label use.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use (ifu) sufficiently address the potential risks.The ifu state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' regarding preparation of the device the ifu state that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment".Regarding the anatomy of the placement site the ifu states: "prior to stent graft deployment, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy".Regarding accessories the ifu states: "prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended".The packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.The ifu states "the safety and effectiveness of the device when placed across an aneurysm or a pseudoaneurysm has not been evaluated".H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent graft placement procedure, the stent graft allegedly would not deploy.It was further reported that the tip of the catheter was allegedly damaged.The procedure was completed using another device.There was no reported patient injury.
|
| |
|