| Model Number |
R560BCS1 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: during preventive maintenance by service technician, it was observed that the bio-console 560 instrument failed the battery voltage test.It measured 11.4v when it should be at least 12.75v.The issue was resolved by replacing the battery,12v,6.5 ah ,certified *2.Preventive maintenance was completed per specifications.Note: the instrument was serviced in the facility by a field service technician.The instrument was not returned to a medtronic facility for service/analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
During preventive maintenance by a service technician this bio-console 560 instrument failed the battery voltage test.It measured 11.4v when it should be at least 12.75v.This was detected during service so there was no patient involvement, so no adverse effect occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that the batteries with lot number 13132513 were installed on 22 feb 2022.Correction d3: mfr name and address updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|