During an oblique lateral lumbar interbody fusion procedure from l2 to l4 with a transforaminal lumbar interbody fusion from l4 to l5, while final tightening, the distal tip fractured off in the right l2 pedicle screw lock screw and could not be retrieved from the patient.
|
|
No device was returned for evaluation and no photographs were provided for review.Additionally, no operative notes were provided for review of usage/technique.Two radiographs were provided; however, based on the images, the complaint was unable to be confirmed as the fractured tip could not bee seen in the lock screw.Information regarding the torque handle used was not provided.Based on the information obtained, the complaint was unable to be confirmed and a root cause was unable to be determined conclusively; however, excessive use of force or technique or material wear/fatigue may have caused or contributed to the reported event.Labeling review: ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." ".Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." ".The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
|