BIOSENSE WEBSTER INC OCTARAY MAPPING CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
|
Back to Search Results |
|
| Catalog Number |
D160904 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a paroxysmal atrial fibrillation procedure with an octaray mapping catheter.As they opened the new octaray mapping catheter, there was a fluff found around one of the splines.Physician did not want to use it inside the patient for safety and infection concerns.Therefore, they changed it to a new one.No patient consequence, multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed on (b)(6)2025.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device number lot 31401479l and no internal action related to the complaint was found during the review.However, if the product is received at a later date, the investigation will be updated as applicable.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|