| Model Number |
8300AB |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Fatigue (1849); Dizziness (2194)
|
| Date of Event |
12/18/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported and through investigation that a patient with a 19mm 8300ab aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 4 months due to stenosis.The patient presented with dizziness and fatigue.The procedure was performed successfully with a 20mm 9755rsl transcatheter valve.Patient left the or with stable vitals and was taken to recovery.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.Through further investigation, it was determined that the root cause of this event was most likely due to patient related factors.
|
| |
|
Search Alerts/Recalls
|