Model Number CNWTT4
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 11/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient had maladaptation to glare.The lens was explanted and replaced with unspecified lens in a secondary procedure.Additional information was requested.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6.And h.11.The product was returned for analysis and the reported complaint could not be confirmed.Iol returned in small plastic bag in sealed white envelope.Solution is dried on the iol.Both haptics are intact.The optic is torn or split cut dividing the iol in three and scratched or marked rejectable.The root cause for the reported complaint could not be determined.The ifu states that some visual effects may be expected due to the superposition of focused and unfocused multiple images.These may include some perception of halos, radial lines around point sources of light(starbursts) under night time conditions, or glare, as well as other visual symptoms.As with other multifocal iols, there is a possibility that visual symptoms may be significant enough that the patient will request explant of the multifocal iol.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21136275
Report Number9612169-2025-00088
Device Sequence Number12641731
Product Code MJP
UDI-Device Identifier00380652468947
UDI-Public(01)00380652468947(17)280408(11)240410(10)25819951
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCNWTT4
Device Catalogue NumberCNWTT4.140
Device Lot Number25819951
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer03/04/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date04/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-