ARROW INTERNATIONAL LLC ARROW AGBA PI MIDLINE 1-L: 4.5FR X 15CM; MIDLINE CATHETER Back to Search Results
Catalog Number CDC-41541-MPKB
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "the wings are breaking off as the sheath is being removed.There was no reported patient harm or consequence." associated complaint 9680794-2025-00060.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).The customer returned one peel-away sheath and product lidstock for analysis.Signs of use in the form of biological material were observed on the sheath extrusion and tabs.Visual inspection revealed that the peel away sheath extrusion was separated directly adjacent to one tab.Microscopic examination confirmed the damage and revealed that the sheath was torn completely down both score lines.It was additionally noted that the tear was partially scalloped which is not the intended appearance of the score lines once torn.The total length of the sheath measured 2 3/4", which is within the specification limits of 2 5/8 - 2 7/8" per the sheath product drawing.The sheath outer and inner diameters could not be accurately measured due to the nature of the damage.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "precaution: avoid tearing sheath at insertion site which opens surrounding tissue creating a gap between catheter and dermis." the report of a broken peel away sheath tab was confirmed through complaint investigation of the returned sample.Visual analysis revealed that the sheath extrusion had separated directly adjacent to one tab.Based on the customer report and the sample received, manufacturing likely caused or contributed to this event.A non-conformance was initiated to further investigate this complaint issue.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that: "the wings are breaking off as the sheath is being removed.There was no reported patient harm or consequence." associated complaints 9680794-2025-00060.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AGBA PI MIDLINE 1-L: 4.5FR X 15CM
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21136475
Report Number9680794-2025-00061
Device Sequence Number6957522
Product Code PND
UDI-Device Identifier10801902205411
UDI-Public(01)10801902205411(11)240918(17)251231(10)33F24J0021(21)
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K161313
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue NumberCDC-41541-MPKB
Device Lot Number33F24J0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2024
Supplement Date Received by Manufacturer02/04/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date02/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-