| Catalog Number |
CDC-41541-MPKB |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that: "the wings are breaking off as the sheath is being removed.There was no reported patient harm or consequence." associated complaint 9680794-2025-00060.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned one peel-away sheath and product lidstock for analysis.Signs of use in the form of biological material were observed on the sheath extrusion and tabs.Visual inspection revealed that the peel away sheath extrusion was separated directly adjacent to one tab.Microscopic examination confirmed the damage and revealed that the sheath was torn completely down both score lines.It was additionally noted that the tear was partially scalloped which is not the intended appearance of the score lines once torn.The total length of the sheath measured 2 3/4", which is within the specification limits of 2 5/8 - 2 7/8" per the sheath product drawing.The sheath outer and inner diameters could not be accurately measured due to the nature of the damage.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "precaution: avoid tearing sheath at insertion site which opens surrounding tissue creating a gap between catheter and dermis." the report of a broken peel away sheath tab was confirmed through complaint investigation of the returned sample.Visual analysis revealed that the sheath extrusion had separated directly adjacent to one tab.Based on the customer report and the sample received, manufacturing likely caused or contributed to this event.A non-conformance was initiated to further investigate this complaint issue.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that: "the wings are breaking off as the sheath is being removed.There was no reported patient harm or consequence." associated complaints 9680794-2025-00060.
|
| |
|
Search Alerts/Recalls
|