| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Child came to hospital for review mapping on (b)(6) 2025, where affected channels were seen.On further enquiry parents informed that child was not responding to sounds since 2 months.Further proceedings are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.No information on possible further steps has been received.
|
| |
|
Event or Problem Description
|
|
Child came to hospital for review mapping on (b)(6) 2025, where affected channels were seen.On further enquiry parents informed that child was not responding to sounds for 2 months.Further proceedings are unknown.
|
| |
|