• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number ACU0T0
Medical Device Problem Codes Failure to Fold (1255); Failure to Prime (1492); Failure to Advance (2524); Difficult to Advance (2920); Failure to Eject (4010)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
A health care professional reported that during surgery the lens would not load or advance due to the blue plastic insert bypassing the lens.Additional information has been requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Additional Manufacturer Narrative
Additional information has been provided in b.2., b.5., d.5., d.10., h.3., h.6., and h.11.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a malfunction.No further reports required.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has received and it states that upon priming, the blue thing inside the cartridge keeps bending, so the lens cannot be pushed straight to properly prime.Surgery able to be completed by the staff but had to open multiple lenses as it kept happening on the new few lenses they have opened.On the 5th lens, scrub nurse managed to have it primed properly.There is no patient contact.No patient impact was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21137951
Report Number1119421-2025-00154
Device Sequence Number6829687
Product Code HQL
UDI-Device Identifier00380652395311
UDI-Public(01)00380652395311(17)270110(11)240112(10)15781645
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACU0T0
Device Lot Number15781645
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/20/2024
Supplement Date Received by Manufacturer01/17/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date02/11/2025
Date Device Manufactured01/12/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC
Patient SexUnknown
-
-