| Model Number |
ECLIPSE CAGE SCREW S, 30MM |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 12/16/2024, it was reported by a sales representative via email that an ar-9400-sbk eclipse speedscap implant system did not get sufficient fixation with the eclipse trunion and cage screw.The case was completed using a long stem.This was discovered during an anatomic total shoulder procedure on (b)(6)2024.Additional information received on 12/30/2024: the part numbers ar-9301-02 and cage screw and ar-9301-45cpc arthrex eclipse trunion were provided.Additionally, the sales representative reported that another ar-9400-sbk eclipse speedscap implant system from a different lot number and an ar-9301-01 arthrex eclipse cage screw was attempted to be used as well but still could not get sufficient fixation.The case was completed using an ar-9100-05m univers apex optifit humeral stem.The case was delayed between thirty to forty minutes; however, additional anesthesia was not needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
|
| |
|
Search Alerts/Recalls
|