Model Number MODULAR POST, 20MM
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 12/17/2024, it was reported by a sales representative via email that an ar-9580-2410 univers revers modular glenoid system augmented baseplate and an ar-9582-20 modular post for augmented mgs baseplate would not engage with the taper.This was discovered during a procedure on (b)(6) 2024.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR POST, 20MM
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21138301
Report Number1220246-2025-09379
Device Sequence Number6807028
Product Code PHX
UDI-Device Identifier00888867341838
UDI-Public(01)00888867341838(11)240301(17)290228(10)1503242317
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K193372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODULAR POST, 20MM
Device Catalogue NumberAR-9582-20
Device Lot Number1503242317
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2024
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date08/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-