| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported by customer that powerpicc solo is leaking, cut/hole in it.Patient who was starting his antibiotics identified the problem before the beginning of the treatment, picc was removed and pivc was inserted for the treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported by customer that powerpicc solo is leaking, cut/hole in it.Patient who was starting his antibiotics identified the problem before the beginning of the treatment, picc was removed and pivc was inserted for the treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a damaged catheter is confirmed; however, the exact cause is unknown.One video of a powerpicc was returned for evaluation.An initial visual observation of the video showed that a clear liquid leaked out from the catheter shaft tubing in the tapered region.When the catheter was brought closer to the camera, a semi-circumferential split could be observed in the catheter tubing; however, there were no distinguishing features in the returned video of the device which could aid in identification of a specific root cause.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|