• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK/DUO-VENT; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK/DUO-VENT; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2R8480
Medical Device Problem Code Filling Problem (1233)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: these events occurred between december 6th, 2024 and december 16th, 2024.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an overinfusion of medication occurred while using an unspecified quantity of clearlink, non-dehp solution sets.The infusions are set for 24 hours and approximately 2l, but the pumps are running out of fluid about 30 minutes early and delivering 50-60 ml more than prescribed.These overinfusions occurred with sigma spectrum pumps during tpn (total parenteral nutrition) treatments.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to h4, h6, and h11: h11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK/DUO-VENT
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
600 mts oeste de entrada
las americas parque industrial
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key21138690
Report Number1416980-2024-07966
Device Sequence Number6807055
Product Code FPA
UDI-Device Identifier00085412565750
UDI-Public(01)00085412565750(10)R24F24062(17)260625
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K153158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/25/2026
Device Model Number2R8480
Device Catalogue Number2R8480
Device Lot NumberR24F24062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2024
Supplement Date Received by Manufacturer02/14/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date02/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPECTRUM INFUSION PUMP; TPN
Patient SexUnknown
-
-