• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 CORAIL AMT SZ11 STD COL; CORAIL IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 CORAIL AMT SZ11 STD COL; CORAIL IMPLANTS : HIP FEMORAL STEM Back to Search Results
Catalog Number L971211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that post-operatively, the patient developed a urinary tract infection which has made them very ill.Infection in the joint has since developed, as surgeon noted patient had been doing well before the above infection.A revision of right tha due to infection following a dair was performed.Patient still had infection and required first stage revision to remove all components, wash out and debridement.Cemented a 52 liner and size 1 c stem placed for approximately 6 weeks, and will then fully revise.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as a valid lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: patient: right side, patient developed a urinary tract infection which has made them very ill.Infection in the joint has since developed as surgeon noted patient had been doing well before the above infection.Patient is considered medically overweight have requested bmi along with additional patient info, awaiting.Event: revision of tha due to infection following a dair (pc: ).Patient still had infection and required first stage revision to remove all components, wash out and debridement.Cemented a 52 liner and size 1 c stem in for approx.6 weeks and will then fully revise.Construct: pinnacle cup (b)(6), dual mobility liner (4313472), pe liner ((b)(6) + femoral head (b)(6), corail stem (b)(6)).The product was not returned to j&j medtech orthopaedics; however, photos were provided for review.(b)(4).The photo investigation revealed foreign substance that appears to be organic material.However nothing indicative of a device nonconformance was found.The observed condition is consistent with the explanation process.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the corail amt sz11 std col would have not contributed to the complained adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT SZ11 STD COL
Common Device Name
CORAIL IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21138860
Report Number1818910-2025-00679
Device Sequence Number12502055
Product Code MEH
UDI-Device Identifier10603295478621
UDI-Public(01)10603295478621
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K213839
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberL971211
Device Lot Number3831184
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/26/2024
Supplement Date Received by Manufacturer01/30/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BI MENTUM ALTRX LNR 47 28.; DLT TS CER HD 12/14 28MM +5.0.; PINN SECTOR W/GRIPTION 54MM.; PINNACLE DM LINER 54_47.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-