• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72077-01
Medical Device Problem Codes Device Alarm System (1012); Application Program Problem (2880)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 12/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
An email complaint was received in which an alarm issue was reported with the abbott diabetes care (adc) device in use with unknown phone with unknown operating system and version.The customer received a signal loss and was not alerted of changes in glucose level.As a result, the customer "fainted for a moment" due to hypoglycemia, requiring apple juice and oral glucose provided by a non-healthcare professional for treatment.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
An investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer reported missing alarms due to signal loss was investigated.After attempting to identify the customer's reported configurations, the information provided in the complaint was insufficient to conduct a comprehensive investigation.The lack of the operating system and app versions restricts the ability to identify potential causes of the customer's reported issue.Therefore, the reported issue is not confirmed due to the inadequate information provided.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.The operating system is unknown so product information provided is for ios.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Common Device Name
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key21139722
Report Number2954323-2025-01207
Device Sequence Number17888063
Product Code QLG
UDI-Device Identifier00357599817008
UDI-Public00357599817008
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K212132
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72077-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2025
Initial Report FDA Received Date01/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-