| Catalog Number |
6260-9-236 |
| Medical Device Problem Codes |
Degraded (1153); Material Erosion (1214); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
It was reported that ¿trunnionosis of hip implant - resulting in broken implant needing to be revised surgically¿.This event was reported to stryker canada by health canada via mandatory hospital reporting: health canada reference #: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that ¿trunnionosis of hip implant - resulting in broken implant needing to be revised surgically¿.This event was reported to stryker canada by health canada via mandatory hospital reporting: health canada reference #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding disassociation and wear involving a metal head was reported.The event was not confirmed.Method & results: material analysis, visual, functional, and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of provided medical records with a clinical consultant indicated "trunnionosis and implant breakage is not confirmed.No documentation was presented to document the event noted in the event description." -device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there has been no other similar event for the lot referenced.Conclusions: a review of provided medical records with a clinical consultant indicated "trunnionosis and implant breakage is not confirmed.No documentation was presented to document the event noted in the event description." the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|