MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM +5 OFFSET; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 36MM +5 OFFSET; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-236
Medical Device Problem Codes Degraded (1153); Material Erosion (1214); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Event or Problem Description
It was reported that ¿trunnionosis of hip implant - resulting in broken implant needing to be revised surgically¿.This event was reported to stryker canada by health canada via mandatory hospital reporting: health canada reference #: (b)(4).
 
Event or Problem Description
It was reported that ¿trunnionosis of hip implant - resulting in broken implant needing to be revised surgically¿.This event was reported to stryker canada by health canada via mandatory hospital reporting: health canada reference #: (b)(4).
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and wear involving a metal head was reported.The event was not confirmed.Method & results: material analysis, visual, functional, and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of provided medical records with a clinical consultant indicated "trunnionosis and implant breakage is not confirmed.No documentation was presented to document the event noted in the event description." -device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there has been no other similar event for the lot referenced.Conclusions: a review of provided medical records with a clinical consultant indicated "trunnionosis and implant breakage is not confirmed.No documentation was presented to document the event noted in the event description." the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM +5 OFFSET
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21143395
Report Number0002249697-2025-00065
Device Sequence Number21233104
Product Code JDI
UDI-Device Identifier07613327032314
UDI-Public07613327032314
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2014
Device Catalogue Number6260-9-236
Device Lot NumberMHLKP1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer04/04/2025
Initial Report FDA Received Date01/13/2025
Supplement Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/25/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-05/07/2018-003R
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-